Product Details


OW-8000A multi-parameter patient monitor is our newest product. The display is 7 inch high resolution color TFT. It is equipped with 4 parameters including ECG, NIBP, SpO2, Resp. These four parameters can be flexible. You can choose different combinations as you need. This monitor is built-in rechargeable lithium battery. The working capacity is 2 hours.


Features
1.  4 Standard parameters: ECG, NIBP, SpO2, Resp
2.  7 inch high resolution color TFT LCD display
3.  Real time S-T segment analysis, pace-maker detection
4.  Indication of incorrect manipulation and malfunction analysis
5.  Patient info input management function
6.  Multi-lead ECG waveforms display in phase
7.  Large volume of tabular and graphic trends information storage and easy to recall
8.  Capture dynamic waveforms
9.  Efficient resistance to interference of defibrillator and HF knife
10. Up to 2 hours working capacity of built-in rechargeable lithium battery
11. Unique accessory management cabinet
12. Networking capacity
13. Option: Temp, EtCO2, Built-in recorder, Cart, Wall shelf.
 
Specification
ECG Specification
Standard 3-lead or 5-lead 3-lead RA,LA,LL,Lead method:I,II,III
5-lead RA,LA,LL,RL,V,Lead method:I,II,III,aVR,aVL,aVF,V
Gain ×250, ×500, ×1000, ×2000
HR Measuring Range Adult:15 ~ 300bpm (beat/minute)
Neonatal baby/children:15 ~ 350 bpm(beat/minute)
HR Accuracy ±1% or ±1bpm,the larger prevails
HR Resolution 1 bpm(beat/minute)
Sensitivity > 200 uV P-P
Differential Input Impedance > 5M ohm
Bandwidth Diagnostic Mode: 0.05~130 Hz
Monitoring Mode: 0.5~40 Hz
Operation Mode: 1~20 Hz
CMRR Diagnostic Mode: > 90 dB
Monitoring Mode: >100 dB
Operation Mode: >100 dB
Electrode Offset Potential ±300 mV
Leakage Current <10 uA
Pacing Pulse Test Test pacing pulse in accordance with the following conditions&colon;
Range&colon; ±2~±700 mV
Width&colon; 0&period;1~2 ms
Rise time&colon; 10~100 us
Pacing pulse inhibition
 
When pacing analysis switch is on&comma; pacing pulse in accordance with the following conditions are restrained&comma; but affection against HR calculation&period;
Range&colon; ±2~±700 mV
Width&colon; 0&period;1~2 ms
Rise time&colon; 10~100 us
Baseline Recovery < 3s After defibrillation
Signal Range ±8mV P-P
Calibrating Signal 1mV&lpar;Peak-to-peak value&rpar;&comma;precision&colon;±5&percnt;
ST Segment Measuring Volume
Measuring Range&colon; -2&period;0mV ~ &plus;2&period;0mV
Measuring precision&colon; Ranging -0&period;8 mV ~ &plus;0&period;8mV&comma; measuring error is ±0&period;02mV or ±10&percnt;&comma;The larger prevails&period;
No definition for other ranges&period;
Resp Specification
Measuring Method RA-LL impedance
RESP Impedance Measuring Range 0&period;3~3&ohm;
Base Impedance Range 200~4000&ohm;
Bandwidth 0&period;1~2&period;5Hz
RR Measurement Range Adult&colon; 0~120BrPM
Children and neonatal baby&colon; 0~150 BrPM
RR Resolution 1 BrPM
RR Precision ±2 BrPM
Asphyxia Alarm 10~40 seconds
NIBP Specification
Measuring Method Pulse wave oscillometry
Work Mode Manual&sol;Automatic&sol;STAT
Measuring Interval of Automatic Measuring Mode 1&comma;2&comma;3&comma;4&comma;5&comma;10&comma;15&comma;30&comma;60&comma;90&comma;120&comma;180&comma;240&comma;480 minute&lpar;s&rpar;
Measuring Time of STAT Mode 5 minutes
PR range 40 - 240 bpm
Measuring Range Adult
Sys&colon; 40~270mmHg
Dia&colon; 10~215mmHg
Mean&colon; 20~235mmHg
Children
Sys&colon; 40~200mmHg
Dia&colon; 10~150mmHg
Mean&colon;20~165mmHg
Neonatal baby
Sys&colon; 40~135mmHg
Dia&colon; 10~100mmHg
Mean&colon; 20~110mmHg
Static Pressure Range 0~300mmHg
Static Pressure Precision ±3mmHg
Pressure precision Max&period; average error&colon; ±5mmHg&semi;
Max&period; standard deviation&colon; 8mmHg
Over voltage protection
 
Adult mode&colon; 300 mmHg
Children mode&colon; 240 mmHg
Neonatal baby mode&colon; 150 mmHg
SpO2 Specification
SpO2 Measuring Range 0~100&percnt;
SpO2 Resolution 1&percnt;
SpO2 Precision 70~100&percnt;&colon; ±2 DIGIT
0&percnt;~69&percnt;&colon; no definition given
PR Measuring Range 20~300bpm
PR Resolution 1bpm
PR Precision ±3bpm
Temp Specification
Temperature Sensor YSI series&comma;CYF series
Channel Quantity 2 Channels
Measuring Range 0~50ºC
Resolution 0&period;1ºC
Precision ±0&period;1ºC&lpar;Excluding Sensor Error&rpar;
IBP Specification
Label ART&comma; PA&comma; CVP&comma; RAP&comma; LAP&comma; ICP&comma; P1&comma; P2
Measuring and alarm range ART&colon; 0~300 mmHg
PA&colon; -6~120 mmHg
CVP&sol;RAP&sol;LAP&sol;ICP&colon; -10~40 mmHg
P1&sol;P2&colon; -10~300 mmHg
Press Sensor Sensitivity 5 uV&sol;V&sol;mmHg
Press Sensor Impedance 300-3000&ohm;
Resolution 1 mmHg
Accuracy ±2&percnt; or±1mmHg&comma; which great
Actualization interval about 1 Sec&period;